Calibration of the sensor


For optimum accuracy, the relation between the measurement signals from the sensor and the tip displacement is measured, as described in this section.


Calibration of the sensor

Before use, each sensor is calibrated to obtain the relationship between the displacement of the sensor tip and the measurement signals produced by the probe. Deviations due to assembly and manufacturing are thus greatly reduced. Calibration of the sensor can be performed on the CMM itself or using the calibration setup.



Setup used to calibrate to Gannen probe.

The relationship between the measurement signal and displacement of the sensor tip is determined using a plane mirror differential laser interferometer setup. A displacement of the measurement mirror results in a displacement of the sensor tip and the resulting measurement signal is recorded. Simultaneously, the displacement of the measurement mirror relative to the reference mirror is measured using a hetrodyne laser interferometer.

The process is repeated at several positions and in different orientations. When the relation between the measurement signal and displacement of the sensor tip for a particular sensor is known, it can be used in measurements.

Gannen 3D TACTILE MEASURING SENSOR FOR CMM METROLOGY
Gannen introduction
Gannen XP, ultra precision
Gannen XM, for MEMS
Gannen Control Unit
Design & support
Measurement results
Probe repeatability
Temperature effects
Uncertainty
Probe calibration
Calibration setup
Frequently Asked Questions

Calibration of the Gannen probe series is performed on a differential laser interferometer setup, developed at the Precision Engineering section at the TU/e, here shown from the top.